ARTHREX, INC. RING CURRETTE,ANKLE ARTHRSCPY,REVSE ANGL; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE
|
Back to Search Results |
|
| Model Number |
RING CURRETTE,ANKLE ARTHRSCPY,REVSE ANGL |
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
01/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 1/30/2023 it was reported by a sales representative via sems that an ar-8655-04 ring currette broke off inside the patient during the ankle arthroscopy/micro fracture repair.The tip of curette broke off inside the joint when trying to remove loose cartilage during the middle of the arthroscopy procedure.The surgeon had to open laterally since he was not able to retrieve the piece inside the joint.They used fluro to located the broken piece but still the surgeon never retrieved the broken piece.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause of this failure is attributed to wear and tear damage incurred over repeatedly use of the device in the field.Manufacturing date 2018.
|
| |
|