Model Number PXB35-08-17-080
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2023
Type of Reportable Event Malfunction
Event or Problem Description
An attempt was made to use a visipro stent to treat a lesion.It was reported that lesion was at the ostium of the right proximal renal artery and evaluated to 80% stenosis. calcium and tortuosity was minimal.Embolic protection was not used.The physician used a 6fr 45 cm non-medtronic sheath to engage the renal.A 0.035 x 260 non-medtronic wire was used to deliver the interventional equipment.   pre dilatation was performed with a 6 x 20 non-medtronic balloon.Post dilatation was performed with a 7x20 non-medtronic balloon.It was reported that stent stent dislodgement occurred during removal following failed delivery.The dislodged stent was removed by capturing in sheath.There was removal difficulties.The delivery catheter caught on stent upon withdrawal.The case was completed with a 6 x 17 x 80 visipro that was post dilated with the 7 x 20 balloon.The case was completed without complication and the patient was discharged the same day without any complaints.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISI-PRO BALLOON-EXPANDABLE STENT SYSTEM
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16404047
Report Number2183870-2023-00060
Device Sequence Number14293873
Product Code NIO
UDI-Device Identifier00643169787834
UDI-Public00643169787834
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P030045
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/29/2025
Device Model NumberPXB35-08-17-080
Device Catalogue NumberPXB35-08-17-080
Device Lot NumberB454673
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2023
Initial Report FDA Received Date02/20/2023
Date Device Manufactured09/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-