| Model Number |
26265 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The patient underwent post transjugular intrahepatic portosystemic shunt for the treatment of large volume clot in the main extrahepatic portal superior mesenteric vein and splenic confluence extending into the right portal vein.A 10x42x135/ 7f reduced profile wallstent rp endoprosthesis was selected to treat the lesion.However, the device was found to be broken proximal to the hub.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: 10x42x135/ 7f unistep plus was received for analysis.The device was returned with the stent in the correct position on the delivery system.A visual and tactile examination identified no issues with the tip of the device.A visual and tactile inspection found the shaft of the device to be separated at approximately 23mm distal of the main t-valve.This type of damage is consistent with excessive force being applied excessive force to the delivery system.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The patient underwent post transjugular intrahepatic portosystemic shunt for the treatment of large volume clot in the main extrahepatic portal superior mesenteric vein and splenic confluence extending into the right portal vein.A 10x42x135/ 7f reduced profile wallstent rp endoprosthesis was selected to treat the lesion.However, the device was found to be broken proximal to the hub.There were no patient complications reported.
|
| |
|
Search Alerts/Recalls
|