| Model Number |
406107 |
| Medical Device Problem Codes |
Material Separation (1562); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An expired fast-cath introducer was used for transvenous pacemaker placement and the venous access detached from the introducer while the device was in the patient.All parts were successfully removed from the patient with no adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis; however, two images were returned.The images appear to show that the side-port tubing detached from the sheath hemostasis body while the sheath was inside the patient.It was also noted that fast-cath hemostasis introducer batch number 7308099 expired on november 30, 2022 which was prior to the reported event date of february 1, 2023.It was confirmed that this product was distributed prior to its expiration date.The cause of the use of expired product is consistent with user error.
|
| |
|