Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having transforaminal lumbar interbody fusion at l3-l4.It was reported that, the cage was large for initial implantation.Broke in half (plastic and titanium separation) before entering the disc space upon malleting. patient presented with a tight l3-4 disc space.Used disc preparation instruments, kerrisons, and drill to adequately prepare site for implantation.Upon malleting implant, the peek and titanium components split apart into two pieces at the connection.The implant split into two distinct and whole pieces and was discarded by staff.The disc space was then filled with a static 10mm height x 22mm length capstone cage.There were no patient symptoms or complications reported.No additional treatment or surgery performed.No further complications reported regarding the event.
|