|
DEPUY SYNTHES PRODUCTS LLC MINIMAL ACCESS ATTACHMENT, STRAIGHT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
|
Back to Search Results |
|
| Model Number |
MA-DRIVER |
| Medical Device Problem Codes |
Complete Loss of Power (4015); Fail-Safe Did Not Operate (4046)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Additional narrative: the actual device was returned for evaluation.The minimal access attachment device was evaluated and the reported condition that the device was not working was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device would run in the locked position.The assignable root cause of this condition was determined to be traced to component failure due to wear.Udi ¿ (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported from india that during service and evaluation, it was determined that the minimal access attachment device had bearing damage and the device would run in the locked position.It was further determined that the device failed pretest for safety assessment.It was noted in the service order that the device was not working.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2022.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|