MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SUMMIT SZ4/5 HI NECK SEGMENT; HIP INSTRUMENTS : FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SUMMIT SZ4/5 HI NECK SEGMENT; HIP INSTRUMENTS : FEMORAL TRIALS Back to Search Results
Model Number 2570-03-250
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Deformation (2976)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/04/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the summit neck trial snap ring was stretched which was preventing the head trials from locking onto the neck trial.When trying to reduce the hip, bipolar head trial would not seat properly onto the neck trial.There was a surgical delay of 3 minutes.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device was received for examination.Therefore, the reported event could not be confirmed.Depuy considers, the investigation closed.Should additional information be received, the information will be reviewed.And the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre) was not possible, because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT SZ4/5 HI NECK SEGMENT
Common Device Name
HIP INSTRUMENTS : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key16409031
Report Number1818910-2023-04136
Device Sequence Number20273452
Product Code LXH
UDI-Device Identifier10603295143109
UDI-Public10603295143109
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2570-03-250
Device Catalogue Number257003250
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2023
Supplement Date Received by Manufacturer03/13/2023
Initial Report FDA Received Date02/21/2023
Supplement Report FDA Received Date03/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age85 YR
Patient SexFemale
-
-