• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH CONNECT FEED STE 1000ML NS; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH CONNECT FEED STE 1000ML NS; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 77100FD
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2023
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the registered nurse (rn) went into the room after the shift report to check on a patient and feeding pump connections.The rn noticed a large air bubble coming from cannister, which measured 99 cm (26 inches) and a second air bubble at the base of the tube towards patient, which measured 12 cm (4 3/4 inches).The rn stopped the feeding pump, obtained a new equipment and feeding bag and started set-up with new equipment.The rn updated the leaving evening rn about the issue found in bag and she did mention that she noticed an air bubble earlier and re-primed the line and told the evening rn of the steps to be taken if encounters the situation again.The original intended procedure for this patient was feeding.There was no harm reported.
 
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Additional Manufacturer Narrative
The device history record review could not be performed since no lot number was provided.The device was received for evaluation and the reported condition has been confirmed.A corrective and preventive action has been initiated to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECT FEED STE 1000ML NS
Common Device Name
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key16410264
Report Number1282497-2023-10738
Device Sequence Number3504700
Product Code LZH
UDI-Device Identifier10884521190825
UDI-Public10884521190825
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 08/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number77100FD
Device Catalogue Number77100FD
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/17/2023
Supplement Date Received by Manufacturer02/17/2023
Initial Report FDA Received Date02/21/2023
Supplement Report FDA Received Date08/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient Age16 YR
Patient SexMale
-
-