Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Events reported via: 2210968-2023-01269, 2210968-2023-01270, 2210968-2023-01271, 2210968-2023-01273.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an aortic aneurysm procedure on (b)(6) 2023 and suture was used.During the procedure, when the surgeon is doing the suturing of the aorta.5 of the sutures was detached from the needle.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 4/5/2023.H6 component code: g07002 no device problem found.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.The returned sample determined that it was received five twenty five unopened samples and five empty labeled winding formers that pertain to the product code 8526h.As per the sampling plan, a visual inspection was performed on thirteen samples, the packets were opened.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand no anomalies were observed during the evaluation.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2023-01269, 2210968-2023-01270, 2210968-2023-01271, 2210968-2023-01273.
|
| |
|