ETHICON INC. PRLNE BLU 30IN 4-0 D/A SH-1; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Model Number 8526H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Events reported via: 2210968-2023-01268, 2210968-2023-01269, 2210968-2023-01270, 2210968-2023-01273.
 
Event or Problem Description
It was reported that a patient underwent an aortic aneurysm procedure on (b)(6) 2023 and suture was used.During the procedure, when the surgeon is doing the suturing of the aorta.5 of the sutures was detached from the needle.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Date sent to the fda: 4/5/2023.H6 component code: g07002 no device problem found.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.The returned sample determined that it was received five twenty five unopened samples and five empty labeled winding formers that pertain to the product code 8526h.As per the sampling plan, a visual inspection was performed on thirteen samples, the packets were opened.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand no anomalies were observed during the evaluation.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2023-01268, 2210968-2023-01269, 2210968-2023-01270, 2210968-2023-01271, and 2210968-2023-01273.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE BLU 30IN 4-0 D/A SH-1
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km. 8.3
san lorenzo
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key16411429
Report Number2210968-2023-01271
Device Sequence Number3289726
Product Code GAW
UDI-Device Identifier10705031016644
UDI-Public10705031016644
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model Number8526H
Device Catalogue Number8526H
Device Lot NumberPLR991
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2023
Supplement Date Received by Manufacturer03/16/2023
Initial Report FDA Received Date02/21/2023
Supplement Report FDA Received Date04/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/10/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-