• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EMPOWR PARTIAL KNEE SYSTEM; EMPOWR PARTIAL KNEETM, TIBIAL INSERT, SIZE 4, 8MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EMPOWR PARTIAL KNEE SYSTEM; EMPOWR PARTIAL KNEETM, TIBIAL INSERT, SIZE 4, 8MM Back to Search Results
Model Number 347-08-704
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/14/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - infection.
 
Additional Manufacturer Narrative
The agent reported "(infection)." the previous surgery and the surgery detailed in this event occurred 2 months apart.This evaluation is limited in scope as limited information was provided to djo surgical - austin for examination.If information regarding cultures identified in the infection, the severity of the infection or any other relevant information is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported device was the source or had a direct connection with the patient's infection.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR PARTIAL KNEE SYSTEM
Common Device Name
EMPOWR PARTIAL KNEETM, TIBIAL INSERT, SIZE 4, 8MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758-5445
MDR Report Key16412841
Report Number1644408-2023-00261
Device Sequence Number5632685
Product Code HSX
UDI-Device Identifier00190446294894
UDI-Public00190446294894
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K191325
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number347-08-704
Device Catalogue Number347-08-704
Device Lot Number166Z1035
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2023
Initial Report FDA Received Date02/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
Patient SexFemale
-
-