| Model Number |
20E |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and verified the reported issue.Stryker replaced the device's user interface pcb assembly to resolve the reported issue.After completing other unrelated repairs, proper device operation was observed through functional and performance testing.The device was returned to the customer for use.Stryker further evaluated the removed user interface pcb assembly and determined that the cause of the reported issue was due to corrupt memory on an integrated circuit, designator u2.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report a non-critical issue with their device.There was no report of patient use associated with the reported event.Upon evaluation of the customer's device, stryker observed that the device¿s display is white.A white display can be indicative of a device lock up condition.As a result, defibrillation therapy may not be available, if needed.
|
| |
|