| Model Number |
TOXO IGM |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer¿s calibration signals were noted to be low.Applicable qc data was not provided.The customer stated qc results were "valid but the bias between points is significant." sample material was requested for investigation but was not available.Based on the information provided, the investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|
Event or Problem Description
|
|
The initial reporter questioned negative results for 2 patient samples tested for elecsys toxo igm (toxo igm) on a cobas 6000 e 601 module compared to 2 other methods.Patient 1 result from the e601 module was 0.335 coi (negative).The result from the siemens method was 1.260 coi (positive).The result from cerba france laboratory was 8.3 coi (positive).Patient 2 result from the e601 module was 0.339 coi (negative).The result from the siemens method was 2.1 coi (positive).The result from cerba france laboratory was 10.4 coi (positive).No questionable results were reported outside of the laboratory.The e601 module serial number was (b)(4).
|
| |
|