| Model Number |
MAJ-2138 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect device was sent to olympus for evaluation.Inspection and testing confirmed the reported issue (broken/unable to connect).A grey clip was broken and detached from the orange connector.The missing grey clip was not returned for evaluation.In addition, the metal connector on the opposite end of the orange connector had several minor scratches on the metal housing.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|
| |
|
Event or Problem Description
|
|
The customer reported the subject device broke and could not be connected in the reprocessing basin.There was no harm or user injury reported due to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 2 years since the subject device was manufactured.Based on the results of the investigation, it is likely the connecting tube could not be connected due to the coming apart of the lock lever by external stress.However, a definitive root cause could not be determined.Olympus will continue to monitor field performance for this device.
|
| |
|