Hold rh 2.22.23
it was reported that surgeon revised a femur fracture on (b)(6) 2023 that failed.A locking screw appears to have possibly been put in a non locking hole and the plate pulled off the bone.One of the 5.0 locking screw was broken, and the shaft fragment remained in the patient.The hardware was placed in (b)(6) 2021.Additionally during this revision, the table top plate bender did not work properly because the screw that secures anvil would not screw in.It is currently unknown if the threads on the screw were damaged or threads on the plate bender.More information will be available after it is inspected post operatively.This report is for one (1) 4.5mm broad lcp plate 14 holes/260mm this is report 1 of 4 for complaint (b)(4).
|
|
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Complainant part is not expected to be returned for manufacturer review/investigation.Date of concomitant therapy is (b)(6) 2023.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device history lot: part #: 226.641, synthes lot #: 6952750, supplier lot #: n/a, release to warehouse date: may 30,2012, manufactured by: synthes elmira.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.
|