• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3371-40QC
Device Problems Over-Sensing (1438); Pacing Asynchronously (1441)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2023
Event Type  malfunction  
Manufacturer Narrative
Session records were provided for review by technical services.Analysis of the session records indicated that the pptwos due to fusion event was sensed by the device.
 
Event Description
It was reported that during a remote follow up, post-paced t-wave oversensing (pptwos) was noted on the device.Abbott technical support was contacted, the session records were analyzed, and the pptwos was suspected to be due to fusion.Reprogramming of the device was recommended.No intervention has been performed at this time.The patient was in stable condition and will continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16420347
MDR Text Key310041688
Report Number2017865-2023-10625
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 02/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCD3371-40QC
Device Lot NumberA000128250
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/14/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-