Model Number BIOPLEX 2200 ANTI-CCP REAGENT PACK
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Bio-rad laboratories has received reports of discrepant results with lot no.301481 from four customers performing lot-to-lot validation against previous lots.Internal testing at bio-rad has demonstrated that lot no.301481 exhibits a reduction in specificity (potential for false positive results).There were no reported adverse event with this lot.However,false positive anti-ccp results could lead to incorrect diagnosis of rheumatoid arthritis and unnecessary medical treatment.
 
Additional Manufacturer Narrative
Bio-rad laboratories has received reports of discrepant results with lot no.301481 from four customers performing lot-to-lot validation against previous lots.Internal testing at bio-rad has demonstrated that lot no.301481 exhibits a reduction in specificity.The bias was observed on the low end of the assay range and none of the false positive results observed exceeded 9 u/ml (3x cut-off).This reduction in specificity may lead to an increase in the number of false positives and could lead to incorrect diagnosis of rheumatoid arthritis and unnecessary medical treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOPLEX 2200 ANTI-CCP REAGENT PACK
Common Device Name
BIOPLEX 2200 ANTI-CCP
Manufacturer (Section D)
BIO-RAD LABORATORIES,INC
4000 alfred nobel dr
hercules 94547
Manufacturer (Section G)
BIO-RAD LABORATORIES, INC
14620 ne north woodinville way
woodinville WA 98072
Manufacturer Contact
ragu muniandy
5500 e second st
benicia, CA 94547
5107415812
MDR Report Key16422988
Report Number1000135116-2023-00001
Device Sequence Number5685251
Product Code NHX
UDI-Device Identifier00847865000857
UDI-Public00847865000857
Combination Product (Y/N)N
PMA/510(K) Number
K093954
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Remedial Action Recall,Notification
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Model NumberBIOPLEX 2200 ANTI-CCP REAGENT PACK
Device Catalogue Number665-3250
Device Lot Number301481
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date02/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/30/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-