| Model Number |
VTICM5_12.6 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The reporter indicated that a 12.6mm vticm5 12.6 implantable collamer lens of -6.00/1.00/89 sphere/cylinder/axis) lens tear/break during loading.Lens tear due to snagging on (b)(6)2022.The lens was not implanted into the patients right eye (od).A replacement lens of the same model/size and similar power was implanted and the problem was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation type 4110: lens work order search- no similar complaint event(s) within associated lots were found.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: h6- type of investigation code: 3331- device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within established process parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim# (b)(4).
|
| |
|