| Model Number |
QD14-G1 |
| Medical Device Problem Codes |
Complete Loss of Power (4015); Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported from switzerland that during service and evaluation, it was determined that the attachment device had liquid damage, heat, and foreign substance/debris/cleaning/sterilization.It was further determined that the device failed pretest for temperature assessment.It was noted in the service order that the device did not work anymore.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The actual device was returned for evaluation.The attachment device was evaluated and the reported condition that the device did not work anymore was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device had heat and liquid damage.The assignable root cause of this condition was determined to be traced to improper reprocessing, which is improper maintenance.Udi ¿ (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|