Concomitant medical products: unknown stapler (unknown stapling device, unk-egiaadapter (unknown egia adapter, evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the device was applied on over indicated tissue.Additionally, the instrument's clamping mechanism was deformed and the anvil was bowed.The product's jaws would not close completely due to a deformed clamping mechanism.It was reported that the device was unable to perform the open/close test prior to use.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur under the following conditions 1.Firing over tissue that is beyond the recommended thickness range.2.Firing with an obstacle incorporated in the jaws.In either of these circumstances, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly, and tissue may not be fully transected.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range thickness for the staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in corresponding selection of staple size.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
According to the reporter, during laparoscopic bypass gastroplasty, when reload was attached to perform the load test, the reload did not open. the doctor also reinserted the load with same results.Another reload was used to resolve the issue in order to complete the case.There was no patient injury.Pli 10 medtronic's initial evaluation of the incident device found the loading unit was partially fired.
|