TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX15RWC W/ 4L RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX15RW40C
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/30/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.Component code: 483 - manifold health effect - impact code: 2645 - no patient involvement health effect - clinical code: 4582 - no clinical signs, symptoms or conditions medical device problem code: 1069 - break investigation findings: 3233 - results pending completion of investigation investigation conclusions: 11 - conclusion not yet available.
 
Event or Problem Description
The user facility reported to terumo cardiovascular that during out of box, the manifold came out damage.*no patient involvement.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on february 23, 2023.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date); g3 (date received by manufacturer) ; g6 (indication that this is a follow-up report) ; h2 (follow-up due to additional information) ; h2 (follow-up due to additional information and device evaluation) ; h3 (device evaluated by manufacturer) ; h6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #3: 3331 - analysis of production records.Investigation findings: 3259 - improper physical structure.Investigation conclusions: 4307 - cause traced to component failure.The returned sample was inspected upon receipt and confirmed that one of the stopcocks on the manifold was slightly bent.A representative retention sample was inspected for damage with no anomalies noted on the device specifically with the manifold.All capiox units are 100% visually inspected at several points in the production and packaging processes.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX15RWC W/ 4L RES
Common Device Name
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key16427652
Report Number1124841-2023-00056
Device Sequence Number16805667
Product Code DTZ
UDI-Device Identifier00699753450790
UDI-Public(01)00699753450790
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K151791
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Model Number3CX*FX15RW40C
Device Catalogue NumberN/A
Device Lot Number2M17
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2023
Supplement Date Received by Manufacturer03/28/2023
Initial Report FDA Received Date02/23/2023
Supplement Report FDA Received Date03/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-