Model Number 1688410110
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was foggy image.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: foggy image probable root cause: cables hdmi cables connectors digital board power supply filter / fuse ac inlet board main board transition board intermittence between transition board and ch connector software camera head (sensor, sensor cable) coupler (dust on optics, scratched optics, broken / shifted lenses, focusing mechanism, endoclamp, zoom ring) problem toggling light source connected monitors leaking poor grounding (e.G camera head connection from cable to transition board.) no/incorrect communication with light source soaking cap disconnection during sterilization electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge cybersecurity attack pendulum ch inertial measurement unit (imu) incident light from surgical scene is reflected by coupler/ch window use error the reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that there was foggy image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIM 4K AUTOCLAVABLE COUPLER, 20MM, C-MOUNT
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key16427909
Report Number0002936485-2023-00150
Device Sequence Number19488345
Product Code GCJ
UDI-Device Identifier07613327562552
UDI-Public07613327562552
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K212511
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1688410110
Device Catalogue Number1688410110
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2023
Supplement Date Received by Manufacturer02/07/2023
Initial Report FDA Received Date02/23/2023
Supplement Report FDA Received Date07/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/01/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-