| Model Number |
115-435 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported left side exchange due to patient¿s concerns with the product.Healthcare professional later reported rupture.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, broken device on radius side assessed as surgical impact opening (shell thickness was within specification) and missing piece of shell assessed as inconclusive.Anxiety-product/procedure: unable to confirm as it is a medical event not related to the device.Additional observations: yellow particles on the surface of the shell.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side exchange due to patient¿s concerns with the product.Healthcare professional later reported rupture.Device has been explanted and replaced.
|
| |
|