| Model Number |
8886623331 |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during an open colostomy closure procedure, when colon anastomosis, the thread broke.In one suture, the thread broke out of the needle and for the other three, it did break during knotting with hands without needle holder.The problem was an instable suture.Needle itself was okay.The device had been opened or removed from the packaging for 40 minutes.To resolve the issue, new suture was used and was put it together with the broken one to finish the suture line.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 8886623331 (8886 6233-31 maxon 4-0 grn 75cm v20x36, lot number: d2d2091y).8886623331 (8886 6233-31 maxon 4-0 grn 75cm v20x36, lot number: d2d2091y).8886623331 (8886 6233-31 maxon 4-0 grn 75cm v20x36, lot number: d2d2091y).Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the suture thread broke and the needle unexpectedly separated from the thread.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|