Model Number 1516-30-407
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Joint Laxity (4526); Unspecified Tissue Injury (4559)
Date of Event 01/01/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2023, the patient underwent the primary surgery via tka for left knee oa with the insert 7mm in question.The surgery was completed successfully without any surgical delay.1 week after the surgery, the insert in question came off.Attempted manual reduction was unsuccessful.On (b)(6) 2023, the revision surgery was done.After initial opening procedure, the following conditions were observed.It was confirmed that the insert had rotated 90°.Her lateral collateral ligament was loose, and the tibial attachment of the patellar ligament was fractured avulsion.The tibial attachment of the patellar ligament was repaired with a high-strength suture.As a result of trial reduction with an insert trial, the tendency to prolapse was subsided with an insert of 12 mm thickness, so a 12 mm thick implant was re-replaced.The revision surgery was completed successfully without any surgical delay.As the cause of this event after primary surgery, it is possible that the flexion gap was loose during the primary surgery, but the selection of the insert with a thin 7mm made it easy to come off.In addition, the patient's delirium progressed after the primary surgery, and it is possible that the insert dislocated due to the unreasonable posture without being able to keep the restricted limb position.The surgeon also left the following comment: although he has never experienced a prolapse when using lcs, he was also concerned about the impact of attune's implant design.No further information is available.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR RP INSRT SZ4 7MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16436407
Report Number1818910-2023-04456
Device Sequence Number15315903
Product Code NJL
UDI-Device Identifier10603295047643
UDI-Public10603295047643
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1516-30-407
Device Catalogue Number151630407
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2023
Supplement Date Received by Manufacturer02/27/2023
Initial Report FDA Received Date02/24/2023
Supplement Report FDA Received Date02/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-