• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON GRIESHABER AG GRIESHABER PNEUMATIC HANDPIECE DSP; UNIT, PHACOFRAGMENTATION, ACCESSORY, HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG GRIESHABER PNEUMATIC HANDPIECE DSP; UNIT, PHACOFRAGMENTATION, ACCESSORY, HANDPIECE Back to Search Results
Catalog Number 725.03
Medical Device Problem Codes Detachment of Device or Device Component (2907); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A customer reported that ophthalmic handpiece tip locking part was not properly engaged and could not be locked when opened.The surgery was performed and completed after replacing the product with another one.The procedure type is unknown.The involved eye and the patient identifier are not available.
 
Additional Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The sample was received with inner blister, outer blister and with cover foil.The sample shows no macroscopic signs of damage.The sample was functionally tested with other device.A disposable forceps was used as an essay.The surgical laser recognized the device.During the investigation, the sample was tested with the foot switch and also with different programs.The device was fully functional.The customer's complaint was not confirmed.A root cause for the customers reported event could not be determined because the returned product met manufacturer¿s specifications; it is not likely that a product malfunction could have contributed to the reported event.No action was taken as the returned sample met specifications for tests performed in relation to the reported event.Complaints are reviewed and monitored at regular intervals for adverse trends.No adverse trends have been observed associated with the reported product and event.No action has been identified for this reported event.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER PNEUMATIC HANDPIECE DSP
Common Device Name
UNIT, PHACOFRAGMENTATION, ACCESSORY, HANDPIECE
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16437348
Report Number3003398873-2023-00029
Device Sequence Number7325790
Product Code HQC
UDI-Device Identifier07612717071391
UDI-Public07612717071391
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K063583
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Catalogue Number725.03
Device Lot Number14EFWX
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/20/2023
Supplement Date Received by Manufacturer03/01/2023
04/17/2023
Initial Report FDA Received Date02/24/2023
Supplement Report FDA Received Date03/29/2023
05/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/28/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-