• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH SIZE 4 ACCOLADE II BROACH; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH SIZE 4 ACCOLADE II BROACH; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 1020-5204
Medical Device Problem Codes Fracture (1260); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/30/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.
 
Event or Problem Description
A size 4 accolade 2 broach was placed in the femur and a calcar planer was used.During the procedure, a rasp handle become unable to attach to the broach.The rasp was pulled out using pliers.When the broach was later checked, the post (protruding part) was damaged, which leading failure to attach the rasp handle to it.
 
Additional Manufacturer Narrative
An event regarding damage involving an accolade broach was reported.The event was confirmed.   method & results:  device evaluation and results: visual inspection noted that the device was returned in used condition.Damage consistent with in-service use was observed on the superior end of the broach.A functional inspection was performed with broach handle (cat 1020-1460, lot: p18n69).There was some difficulty assembling the broach to the handle without fully locking in.The event is confirmed.Damage consistent with in-service use clinician review: no medical records were received for review with a clinical consultant.Device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: visual inspection noted that the device was returned in used condition.Damage consistent with in-service use was observed on the superior end of the broach.A functional inspection was performed with broach handle (cat 1020-1460, lot: p18n69).There was some difficulty assembling the broach to the handle without fully locking in.The event is confirmed.Damage consistent with in-service use.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
A size 4 accolade 2 broach was placed in the femur and a calcar planer was used.During the procedure, a rasp handle become unable to attach to the broach.The rasp was pulled out using pliers.When the broach was later checked, the post (protruding part) was damaged, which leading failure to attach the rasp handle to it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIZE 4 ACCOLADE II BROACH
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16438038
Report Number0002249697-2023-00188
Device Sequence Number3151153
Product Code LPH
UDI-Device Identifier07613327252620
UDI-Public07613327252620
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153345
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1020-5204
Device Catalogue Number1020-5204
Device Lot Number143250
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/30/2023
Supplement Date Received by Manufacturer03/17/2023
Initial Report FDA Received Date02/24/2023
Supplement Report FDA Received Date04/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/06/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-