• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9600TFX23J
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Ischemia Stroke (4418)
Event Date 06/08/2022
Event Type  Injury  
Manufacturer Narrative
Per the instructions for use (ifu), permanent or transient neurological events such as transient ischemic attack (tia) and stroke are known potential adverse events associated with the transcatheter heart valve (thv) procedure and the use of the edwards thv devices.According to the literature review, and as documented in a clinical technical summary written by edwards lifesciences, stroke is recognized in the literature as a well-known complication in a small number of patients undergoing thv.Risk factors correlating with several patient co-morbidities have been identified.Although in many cases the root cause of the event is unable to be determined, strokes during thv are undoubtedly multifactorial, the dominant etiology likely being intra-procedure embolic events.A transcranial doppler study during thv demonstrated that most procedural embolic events occurred during balloon valvuloplasty, manipulation of catheters across the aortic valve, and valve implantation.An analysis in patients undergoing valve surgery revealed four baseline characteristics and two procedural events that were associated with early post-procedure stroke: female sex, ef < 30%, diabetes, age older than 70 years, bypass procedure time> 120 min, and calcification of the ascending aorta.Predictors of late stroke have included female sex, age older than 75 years, atrial fibrillation, and a history of or current smoking.There were no significant differences in the frequency of late strokes between thv and aortic valve replacement (avr) patients.After thv, there appears to be a more considerable proportion of early strokes occurring < 24 h post-procedure, but thv patients with multiple co-morbidities are probably at higher risk of both early and late strokes.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results are inconclusive as patient and procedural factors were not provided; however, the event could be related to the mechanisms described above.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.
 
Event Description
As reported through the japanese tavi registry, after a transfemoral aortic valve replacement (tavr) procedure with a 23mm sapien 3 valve in the aortic position, on postoperative day (pod) 5, no obvious paralysis was observed, but mini-mental state examination (mmse) score after tavr procedure dropped from 27 to 13 points.Magnetic resonance image (mri) showed a new cerebral infarction.The patient was discharged on pod 8 after continuing medication without any symptoms.Per the medical opinion, the event relationship to both the valve and the tavr procedure was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key16438525
MDR Text Key310185700
Report Number2015691-2023-11068
Device Sequence Number1
Product Code NPT
UDI-Device Identifier00690103195033
UDI-Public(01)00690103195033(17)231018
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/18/2023
Device Model Number9600TFX23J
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Hospitalization;
Patient Age87 YR
Patient SexMale
-
-