| Model Number |
670175 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a cuff leak was discovered during pre-test after reprocessing when product was about to be reinstalled into patient.No injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer and h6.Health impact, and evaluation codes: updated.One device was received in used condition without the original packaging.Visual inspection detected that the cuff had damage (tear).A leak test was performed to confirm the defect in the sample, the sample failed the test confirming the complaint and a root cause was due to improper use of the product.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.
|
| |
|