| Catalog Number |
07.02117.001 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a vitality probe fractured intra-operatively.The fractured pieces were retrieved and a different instrument was used to complete the procedure.There was a 5 minute delay, but no impact to the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that a vitality probe fractured intra-operatively.The fractured pieces were retrieved and a different instrument was used to complete the procedure.There was a 5 minute delay, but no impact to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in g1 and h3.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed that one of the tips of the probes was fractured off.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.This event could also be attributed to off-axis or excessive forces applied during use.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a vitality probe fractured intra-operatively.The fractured pieces were retrieved and a different instrument was used to complete the procedure.There was a 5 minute delay, but no impact to the patient.
|
| |
|