| Model Number |
33683206 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Wound Dehiscence (1154); Implant Pain (4561)
|
| Date of Event |
01/30/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.Device remains implanted in patient.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a total ankle replacement.The patient presented with wound break down and pain.Upon a ct scan it looked like there was loosening but once it was cleaned out, the implant was stable aseptic, cultured during wash out and nothing grew.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.An assessment could not be made due to the poor quality of the ct scans.There are many clinical factors that can affect the results of any surgery, such as surgical technique, pre-operative and post-operative care, the implant, patient pathology and daily activity.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a total ankle replacement.The patient presented with wound break down and pain.Upon a ct scan it looked like there was loosening but once it was cleaned out, the implant was stable aseptic, cultured during wash out and nothing grew.
|
| |
|
Search Alerts/Recalls
|