ZIMMER SWITZERLAND MANUFACTURING GMBH BIOLOX DELTA, CERAMIC FEMORAL HEAD, L, 40/+3.5, TAPER 12/14; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER
|
Back to Search Results |
|
| Model Number |
N/A |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
03/21/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: medical devices: continuum tm cluster holes porous; item# 00-8757-056-01; lot# 61788312 biolox delta ceramic taper liner, size kk / 40 i.D.For use with 56 mm o.D.Size kk shell; item# 00-8775-012-40; lot# 2523977 ml taper, plasma sprayed, calcicoat, extended offset; item# 6577111120; lot# 61574077 multiple mdr reports were filed for this event, please see associated reports: 0009613350 - 2023 - 00079.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a total hip arthroplasty and approximately 9 years later had a dislocation of the right hip (operated side).Relocation was performed under sedation in a&e.Due diligence is in progress for this complaint; to date no additional information or product has been received.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) this follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h1, h2, h3, h6, h10.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirms no abnormalities or deviations.Devices are used for treatment.Medical records were not provided.With the available information, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|