BIOMET UK LTD. OXF ANAT BRG LT SM SIZE 4 PMA; KNEE PROSTHESIS Back to Search Results
Model Number 159541
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Foreign: (b)(6).Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported the inlay (bearing) fractured after cementless knee implantation in 2008, without adequate trauma.No remedial action has been reported.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.D10: associated products: item number: 154925, item name: oxford ph3 cementless fem sz s, item lot: unknown item number: 154915, item name:oxf ph3 cementless tib sz b lm, item lot: unknown no product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of manufacturing records cannot be performed without product lot number being provided.Review of complaint history identified additional similar complaints for the reported item.As the lot number is unknown, an additional review could not be performed.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.Device is used for treatment.Radiographs were provided and reviewed by a health care radiologist.Impressions are narrowing of the medial compartment noted on images acquired at and after 08/09/2021 which is suggestive of nonspecific disruption or wear of the polyethylene lining in this region.This results in narrowing of the medial compartment joint space.Medical records were not provided.With the available information, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the patient underwent a bearing revision due to fracture without adequate trauma, approximately 15 years post-op.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF ANAT BRG LT SM SIZE 4 PMA
Common Device Name
KNEE PROSTHESIS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
.
warsaw, IN 46582
5745273773
MDR Report Key16442278
Report Number3002806535-2023-00048
Device Sequence Number5417018
Product Code NRA
UDI-Device Identifier05019279785872
UDI-Public(01)05019279785872
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P010014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number159541
Device Catalogue Number159541
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2023
Supplement Date Received by Manufacturer05/04/2023
Initial Report FDA Received Date02/27/2023
Supplement Report FDA Received Date05/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Required Intervention; Other; Hospitalization;
Patient Age80 YR
Patient SexFemale
-
-