• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SWITZERLAND MANUFACTURING GMBH PROXIMAL HUMERUS, RIGHT, LONG, 8.5X240MM; ROD, FIXATION, INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SWITZERLAND MANUFACTURING GMBH PROXIMAL HUMERUS, RIGHT, LONG, 8.5X240MM; ROD, FIXATION, INTRAMEDULLARY Back to Search Results
Model Number N/A
Medical Device Problem Codes Patient Device Interaction Problem (4001); Migration (4003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical products:¿ blunt tip screw, 4x40mm; item#47248604040; lot#3078169, blunt tip screw, 4x40mm; item#47248604040; lot#3082108, torque limiting handle, a.5; item#27923; lot#unknown.Japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see the associated reports: 0009613350 -2023 -00086, 0009613350 - 2023 -00089.
 
Event or Problem Description
It was reported that the patient underwent revision surgery due to migration.Approximately one and half months post-op.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).D10: blunt tip screw, 4x40mm; item#47248604040; lot#3078169.Blunt tip screw, 4x40mm; item#47248604040; lot#3082108.Affixus screw trauma impl win gen; item#unknown; lot#unknown.Torque limiting handle, a.5; item#27923; lot#unknown.This follow-up report is being submitted to relay additional and/or corrected information.Two blunt tip screws were returned for investigation.The other products involved in the reported event were not returned.The two blunt tip screws were visually inspected and on both of them it is possible to see that the threaded parts are partially flattened.The heads of the blunt tip screws appear inconspicuous.A review of the device manufacturing records (for the nail and 2 blunt screws) confirmed no abnormalities or deviations.The review of the device manufacturing records for the unknown screw could not be performed.Devices used for treatment.X-ray images of the affected humerus are available.The x-rays were reviewed by a radiologist: two views of the right humerus proximally demonstrate an intramedullary rod with oblique fracture line still seen along the proximal to mid humeral diaphysis.Three proximal interlocking screws are present, all of which appear to have backed out somewhat most prominently involving the most proximal screw.No surgical reports were provided.Based on the investigation it could be assumed that further possible contributing factors to the migration of the screw might be multifactorial related to either patient condition, behavior or implantation procedure.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.An additional deeper investigation was performed which identified the design limitation of the corelock mechanism of the affixus nail as a potential contributing factor.As part of the deeper investigation, a scientific literature search was conducted and a systematic review of the outcome of intramedullary nailing for acute proximal humerus fracture showed that screw migration and perforation into the joint is the second most common complication following secondary loss of reduction.In addition, scientific literature of competitor and predicate devices were assessed.This review has shown that the affixus natural nail proximal humeral system performs better and/or in equal range in comparison with legally marketed similar devices.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw.Multiple mdr reports were filed for this event, please see the associated reports: 0009613350 - 2023 - 00086, 0009613350 - 2023 - 00089, 0009613350 - 2023 - 00198.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROXIMAL HUMERUS, RIGHT, LONG, 8.5X240MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY
Manufacturer (Section D)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER SWITZERLAND MANUFACTURING GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key16442292
Report Number0009613350-2023-00083
Device Sequence Number20219869
Product Code HSB
UDI-Device Identifier00889024505964
UDI-Public(01)00889024505964(17)311014(10)3087021
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K181827
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number47-2496-240-08
Device Lot Number3087021
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2023
Supplement Date Received by Manufacturer04/20/2023
Initial Report FDA Received Date02/27/2023
Supplement Report FDA Received Date05/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexPrefer Not To Disclose
-
-