• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS VOLISTA STANDOP; LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS VOLISTA STANDOP; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number ARD267900100A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2023
Type of Reportable Event Malfunction
Event or Problem Description
On 20th february 2023 getinge became aware of an issue with one of surgical lights - volista standop.As it was stated, one of the six fixation screws holding the main tube was loosen.There was no injury reported, however, we decided to report the issue in abundance of caution as defect of the screws involved in this complaint may lead to the detachment of the whole configuration and serious injury in case of event reoccurrence.
 
Additional Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.Device not returned to manufacturer.
 
Event or Problem Description
Manufacturer's reference number (b)(4).
 
Additional Manufacturer Narrative
Getinge became aware of an issue with one of surgical lights - volista standop.As it was stated, one of the six fixation screws holding the main tube was loosened.There was no injury reported, however, we decided to report the issue in abundance of caution as defect of the screws involved in this complaint may lead to the detachment of the whole configuration and serious injury in case of event reoccurrence.According to the information provided by getinge technician, the loosened screw (ard367270555) was replaced.It was established that when the event occurred, the surgical light did not meet its specification due to fixation screws holding the device main tube being loose and, in this way, contributed to the event.Information gathered during the investigation indicates that the device was not being used for patient treatment when the event occurred.A review of received customer product complaints related to investigated issues, revealed that there were no injuries to a user nor to a patient or operator when this particular malfunction occurred.Comparing the number of claimed devices to the number of sold devices worldwide, we can assume that the failure ratio of the issue investigated herein is very low.During the analysis, the subject matter expert at the manufacturing site determined the following to have likely occurred.A loosening of the screw or screws, where the probable causes of loosening corresponds to an improper initial tightening during the installation or a maintenance not in accordance with the manufacturer's recommendations (installation manual ni 01784 en 9, page 30).In case of rupture, the failure of the screws is related to fatigue stresses due to moderate shear overload.All facies have a fatigue failure zone more or less extensive.Initiated at the bottom of the thread, the repeated low mechanical stresses caused the breakage of a first screw.The rupture of the other screws corresponds to progressive propagation of the shear effect.To prevent any incident the yearly preventive maintenance program, mentions to check the tightening of fixation screws on the suspension tube (01762 en 08, page 21, 215).In april 2019 getinge transmitted the service bulletin 98a on getinge online reminding to carry out preventive maintenance and wearing parts replacement to ensure the device safety, the service bulletin 98a mentions to replace the suspension fixing screws every 6 years during the preventive maintenance.We believe the related devices are performing correctly in the market.We also believe that if the manufacturer¿s recommendation had been followed the incident could have been avoided.Getinge shall continue to monitor for any further events of this nature and do not propose any further action at this time.The correction of d4 catalog #, d4 version of model #, d4 serial # and e1b event site name fields deems required.This is based on the internal evaluation.Previous d4 version of model # ard267900100c.Corrected d4 version of model # ard267900100a.Previous d4 catalog # ard267900100c.Corrected d4 catalog # ard568831963.Previous d4 serial # (b)(6).Corrected d4 serial # (b)(6) /(b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOLISTA STANDOP
Common Device Name
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key16442507
Report Number9710055-2023-00134
Device Sequence Number5531048
Product Code FSY
UDI-Device Identifier03700712406004
UDI-Public(01)03700712406004
Combination Product (Y/N)N
Initial Reporter CountryHK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARD267900100A
Device Catalogue NumberARD568831963
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2023
Supplement Date Received by Manufacturer10/12/2023
Initial Report FDA Received Date02/27/2023
Supplement Report FDA Received Date10/27/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/27/2019
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-