MEDTRONIC HEART VALVES DIVISION EVOLUT FX DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
Model Number D-EVOLUTFX-2329 |
Device Problems
Difficult to Advance (2920); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/02/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Product analysis: the device was discarded; therefore, no product analysis can be performed.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, via bilateral femoral access obtained by the interventional cardiologist (ic), the valve was loaded.Upon insertion of a large bore sheath, it was noted the ic had difficulty advancing the delivery catheter system.A contrast injection showed vascular complications at the insertion site.Subsequently, due to the vascular complications and the need to repair the femoral artery, the procedure was aborted.It was noted the patient would be scheduled at a later date for the valve implant procedure to be performed.No additional adverse patient effects were reported.
|
|
Event Description
|
Additional information was received that prior to the insertion of the delivery catheter system (dcs), while the physician was exchanging introducer sheaths, a guidewire dissection occurred.The dcs was never inserted into the patient.Subsequently a covered stent was placed resolving the dissection.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Medtronic received additional information that changed the event description and device relatedness of this previously reported event.There is no evidence to suggest the device caused or contributed to a death or serious injury.Updated data: a4 - patient weight b5 - event description b7 - relevant history h6 - device code and conclusion code medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|