• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE COOK CELECT FEMORAL AND JUGULAR VENA CAVA FILTER; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE COOK CELECT FEMORAL AND JUGULAR VENA CAVA FILTER; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G48378
Medical Device Problem Codes Structural Problem (2506); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Chest Pain (1776); Dyspnea (1816); Internal Organ Perforation (1987); Perforation of Vessels (2135); Ambulation or Postural Difficulties (2544); Insufficient Information (4580)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Occupation: non-healthcare professional.Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown, however, the alleged celect is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It is alleged that the patient received a celect inferior vena cava (ivc) filter on (b)(6) 2012, and the patient was injured without further explanation.Hospital and medical records have been requested, but not yet provided.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.The following fields were updated per additional information received: investigation the following allegations have been investigated: organ/vena cava perforation.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma.Catalog number and lot number are unknown; however, the alleged filter is manufactured and inspected according to controls.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
(b)(6) 2020, per a report from computed tomography; ¿ivc filter with one leg extending through the walls of the ivc and into the wall of the third part of the duodenum.No associated pathology is apparent".
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.The following fields were updated per additional information received: a4, b5, b6, b7, d1, d4, g4, h4, h6.Investigation the following allegations have been investigated: dyspnea, chest pains, limited activity.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Unknown if the reported dyspnea, chest pains, and limited activity are directly related to the filter and unable to identify a corresponding failure mode at this point in time.(b)(4) devices in lot.No relevant notes on work order.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Patient allegedly received an implant via the right internal jugular vein on (b)(6) 2012 due to pulmonary embolism (pe).The patient alleges organ perforation.The patient additionally alleges difficulty breathing, chest pains, limited activity, and shortness of breath (sob).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT FEMORAL AND JUGULAR VENA CAVA FILTER
Common Device Name
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
lissi walmann
sandet 6, dk-4632
bjaeverskov 
MDR Report Key16443151
Report Number3002808486-2023-00052
Device Sequence Number5835122
Product Code DTK
UDI-Device Identifier00827002483781
UDI-Public(01)00827002483781(17)140319(10)E2891269
Combination Product (Y/N)N
PMA/510(K) Number
K073374
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/19/2014
Device Model NumberG48378
Device Catalogue NumberIGTCFS-65-UNI-CELECT-PERM
Device Lot NumberE2891269
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/22/2023
Supplement Date Received by Manufacturer03/01/2023
05/08/2023
Initial Report FDA Received Date02/27/2023
Supplement Report FDA Received Date03/20/2023
05/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/01/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient SexFemale
Patient Weight91 KG
-
-