Model Number 1458Q/86
Device Problems Failure to Capture (1081); Capturing Problem (2891)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2023
Event Type  Injury  
Event Description
Related manufacturer reference number: 2017865-2023-10964.It was reported episodes of inappropriate automatic mode switch due to noise was observed on the right atrial (ra) lead, and high capture threshold was observed on the left ventricular (lv) lead.The noise was not reproducible with isometrics.The leads were explanted.The patient was stable.
 
Event Description
Related mfr report number: 2017865-2023-10964.Additional information received notes insulation failure on the atrial lead and loss of capture on the left ventricular (lv) lead.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16443488
MDR Text Key310247784
Report Number2017865-2023-10965
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2023
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000100162
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/19/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA RV LEAD; QUADRA ASSURA ICD
Patient Outcome(s) Required Intervention;
Patient Age67 YR
Patient SexFemale
Patient Weight131 KG
Patient RaceBlack Or African American
-
-