| Model Number |
35615 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that radioopaque marker location was not in its normal position.The non stenosed target lesion was located in mildly tortuous and non calcified posterior back.A.038 j accustick ii was selected for use.During the procedure, it was noted that the radio plaque marker was not in its normal position.It was displaced and farther up the introducer than normal.The device was removed over the wire and the procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the device was returned for evaluation.The accustick device was returned and it was observed that the ro marker was out of its location in the device middle section since it moves freely.Additionally, the accustick device was bent and had the distal tip was broken.No more damages were observed.The device was inspected under microscope and it was observed that the distal tip was broken, this damage is located next to the ro marker position.It's important to mention that evidence of correct ro marker colocation was observed at device distal end section.The device was measured in order to confirm the ro marker position; and the ro marker was located approximately at 3.5cm from the distal tip, and the specification is: 6.50mm0.50mm.The ro market was measured by outer diameter and the results were 0.0823in, and the specification is: 0.08200.0005.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that radioopaque marker location was not in its normal position.The non stenosed target lesion was located in mildly tortuous and non calcified posterior back.A.038 j accustick ii was selected for use.During the procedure, it was noted that the radio plaque marker was not in its normal position.It was displaced and farther up the introducer than normal.The device was removed over the wire and the procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Search Alerts/Recalls
|