|
H10: h3, h6: the reported device was received for evaluation.A visual inspection revealed the device is not in its original packaging.The insertion device was returned with the sutures and anchor still installed.The anchor is fractured at the suture window.There is debris on the returned items.Based on the condition of the product material found during visual inspection, it was determined the device damage was caused by the application of improper/excessive force.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found the raw material strength requirements and storage requirements specified and each lot must be accompanied by a material certificate of analysis.A clinical review states a single undated, unlabeled image that shown a broken screw was provided that confirms the breakage.The root cause has been associated with unintended use of the device.Factors, which could have contributed to the complaint event, include an application of unintended inappropriate or excessive force to the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|