| Model Number |
BIO-COMPRESSION SCREW 3 X 26MM |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2023, it was reported by a sales representative via sems that an ar-5025b-22bio-compression screw,3.0mm by 22mm, and an ar-5025b-26 bio-compression screw, 3.0mm by 26mm, snapped in half during insertion.Case was completed by using a product from another manufacturer.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to improper bone preparation, misaligned insertion and/or prying/leveraging during use.
|
| |
|
Search Alerts/Recalls
|