• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BIO-COMPRESSION SCREW 3 X 26MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BIO-COMPRESSION SCREW 3 X 26MM; SCREW, FIXATION, BONE Back to Search Results
Model Number BIO-COMPRESSION SCREW 3 X 26MM
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/07/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On (b)(6) 2023, it was reported by a sales representative via sems that an ar-5025b-22bio-compression screw,3.0mm by 22mm, and an ar-5025b-26 bio-compression screw, 3.0mm by 26mm, snapped in half during insertion.Case was completed by using a product from another manufacturer.
 
Additional Manufacturer Narrative
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to improper bone preparation, misaligned insertion and/or prying/leveraging during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-COMPRESSION SCREW 3 X 26MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key16444477
Report Number1220246-2023-06350
Device Sequence Number14852893
Product Code HWC
UDI-Device Identifier00888867034730
UDI-Public00888867034730
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K060478
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 09/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model NumberBIO-COMPRESSION SCREW 3 X 26MM
Device Catalogue NumberAR-5025B-26
Device Lot Number15014056
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2023
Supplement Date Received by Manufacturer08/12/2024
Initial Report FDA Received Date02/27/2023
Supplement Report FDA Received Date09/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-