| Model Number |
406948 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/20/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During the atrial fibrillation procedure, the introducer got stuck in the femoral vein and broke apart.The lock tore and there was no resistance.The half of the introducer that was out of the patient was frayed on the end.The other half of the introducer was still in the patient's body, requiring vascular surgery for removal.A non-abbott merit medical snare was used and the other half of the introducer was successfully extracted.The introducer had several deformations due to the use of the one snare.There were no complications during the procedure and it was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
One 8f fast-cath introducer sheath was received for evaluation.The sheath tubing had been torn and stretched into two sections.There were multiple bends and kinks throughout the distal and proximal sheath tubing sections.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the entrapment and sheath tubing damage remains unknown.
|
| |
|