| Model Number |
1010 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device was thoroughly inspected and analyzed.Visual inspection identified no anomalies.The device was able to be interrogated and a memory download was performed successfully.The device was then exposed to simulated heart load conditions, and the defibrillation and sensing functions were tested.The device operated appropriately, according to its performance specifications with no out of range measurements or interruptions in therapy output.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.Representatives from our quality assurance, manufacturing, and design engineering groups are actively investigating this behavior.Our ongoing investigation efforts are currently focused on understanding the root cause to mitigate the potential for setscrew issues when disconnecting the electrode from these s-icds.Information gained through these efforts will be utilized to implement appropriate preventive action(s), as necessary.
|
| |
|
Event or Problem Description
|
|
It was reported that during routine replacement of this subcutaneous implantable cardioverter defibrillator (s-icd), difficulty was encountered with loosening the set screw to remove the electrode from the device header.Multiple torque wrenches and techniques were attempted, however, the set screw was unable to be loosened.The physician then tugged on the electrode and it came out of the device header.The device was successfully explanted and replaced.No adverse patient effects were reported.The product has been received for analysis.
|
| |
|
Search Alerts/Recalls
|