BOSTON SCIENTIFIC CORPORATION SQ-RX PULSE GENERATOR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 1010
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/02/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned device was thoroughly inspected and analyzed.Visual inspection identified no anomalies.The device was able to be interrogated and a memory download was performed successfully.The device was then exposed to simulated heart load conditions, and the defibrillation and sensing functions were tested.The device operated appropriately, according to its performance specifications with no out of range measurements or interruptions in therapy output.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.Representatives from our quality assurance, manufacturing, and design engineering groups are actively investigating this behavior.Our ongoing investigation efforts are currently focused on understanding the root cause to mitigate the potential for setscrew issues when disconnecting the electrode from these s-icds.Information gained through these efforts will be utilized to implement appropriate preventive action(s), as necessary.
 
Event or Problem Description
It was reported that during routine replacement of this subcutaneous implantable cardioverter defibrillator (s-icd), difficulty was encountered with loosening the set screw to remove the electrode from the device header.Multiple torque wrenches and techniques were attempted, however, the set screw was unable to be loosened.The physician then tugged on the electrode and it came out of the device header.The device was successfully explanted and replaced.No adverse patient effects were reported.The product has been received for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SQ-RX PULSE GENERATOR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
229 avenida fabricante
san clemente CA 92672
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16446517
Report Number2124215-2023-08099
Device Sequence Number14921752
Product Code LWS
UDI-Device Identifier00802526521300
UDI-Public00802526521300
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P110042/S000
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date02/28/2015
Device Model Number1010
Device Catalogue Number1010
Device Lot NumberA010205
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/17/2023
Initial Report FDA Received Date02/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
-
-