| Model Number |
03R89-62 |
| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Updated postal code (b)(6).An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer observed that hiv results auto released from aniq ams software that connected to alinity m processing module.The customer has range setup in ams software that hold all hiv results if >50 for manual verification.That range was setup in ams software at the time of installation and that validated by the customer.On 05feb2023 all hiv results started auto released without holding if results >50.The abbott fsr found that samples are partially validated because results are not being copied to the lis hiv dummy test code causing samples to remain on ¿to validate¿ list and test type was changed to textual a day before by abbott remotely.The abbott fsr restarted the services, and the results were copied to the dummy test as samples were closing properly.There is no information regarding any discrepant results reported.No further impact to patient management reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The abbott specialist performed technical investigations, revealed the issue was a configuration error performed by the abbott specialist that led to a hiv workflow and sample management.A review of 12 months of complaint data did not identify any adverse trends or any non-statistical trends or any atypical complaint activity associated with this described issue.A review of tracking and trending of the aliniq found no additional complaints replated to the complaint issue.Therefore, no trends were identified.A review of the manufacturing documentation did not identify any issues associated with the complaint issue.A review of the labeling addresses the customer¿s issue: aliniq ams configurator user manual en 3.01 rev0 provides adequate information about how to configure the test type in aliniq ams and the validation method in the analyzer section.Based on the information within the complaint, a malfunction was identified: a human (abbott specialist) error in the aliniq ams configuration was the cause of the reported issue which caused a failure in the device to perform as intended at the customer site.No deficiencies were identified in the aliniq ams middleware.
|
| |
|
Search Alerts/Recalls
|