| Model Number |
S-65-040-120-P6 |
| Medical Device Problem Codes |
Activation Failure (3270); Migration (4003)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported activation/deployment failure appears to be related to circumstances of the procedure as it is likely that the during stent deployment the introducer sheath was too close to the proximal end of the stent; thus resulting in the stent to partially deploy into the introducer sheath.As reported, when the sheath was pulled the stent moved into the common femoral/external iliac artery and remains there; thus resulting in the reported migration.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat the 6.5 mm proximal superficial femoral artery with no calcification or tortuosity.The vessel was prepared with a 7x40 mm balloon.Atherectomy was not used.The 6.5x40 mm supera self expanding stent (ses) partially (about 5 mm) deployed in the sheath.When the sheath was pulled, the stent moved into the common femoral/external iliac artery and remains there.This is an unintended location.The target lesion was not treated.
|
| |
|