Model Number 1458Q/86
Device Problems Failure to Capture (1081); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2023
Event Type  Injury  
Event Description
It was reported that a patient presented in-clinic.Examination via x-ray imaging confirmed dislodgement of the left ventricular lead, leading to loss of capture.The lead was explanted and replaced on (b)(6) 2023.No patient consequences were reported.
 
Manufacturer Narrative
Correction: h6: codes should reflect product not returned.
 
Manufacturer Narrative
The reported events were for lead dislodgement and failure to capture.As received, a complete lead was returned in one piece.Visual inspection of the lead found no anomalies.The s-curve hump height was measured to be within product specification.The reported event for failure to capture was not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16448839
MDR Text Key310317277
Report Number2017865-2023-11014
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 05/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000123567
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/05/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATATACHY.; GALLANT HF.; TENDRIL LEAD.
Patient Outcome(s) Required Intervention;
Patient Age70 YR
Patient SexMale
-
-