Model Number CDVRA500Q
Device Problem Failure to Interrogate (1332)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2023
Event Type  malfunction  
Event Description
It was reported that a patient presented in-clinic with loss of bluetooth low energy telemetry noted on the patient's implantable cardioverter defibrillator.No intervention or adverse patient consequences were reported.
 
Manufacturer Narrative
The reported event of an inability to communicate via remote monitoring was confirmed.The issue was resolved by bring the patient into clinic for device interrogation.The results of the software analysis are inconclusive since no additional information was obtained to investigate.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT VR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16448931
MDR Text Key310331944
Report Number2017865-2023-11015
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067031990
UDI-Public05415067031990
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2023
Device Model NumberCDVRA500Q
Device Catalogue NumberCDVRA500Q
Device Lot NumberS000082982
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA TACHY
Patient Age53 YR
Patient SexFemale
-
-