Model Number 2738
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 02/01/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the device allegedly caused pain and a burn to the patient during treatment.
 
Additional Manufacturer Narrative
It was reported that the device allegedly caused pain and a burn to the patient during treatment.The degree of burn was not reported, it was reported the patient was seeing a dermatologist for treatment.Enovis performed an inspection on the actual device returned.The unit passed functional and safety testing.The product met manufacturer specifications.If additional reporting of this event becomes available it will be provided as a supplemental to this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELECT TRANSPORT COMBO PKG US STD
Common Device Name
STIMULATOR, ULTRASOUND AND MUSCLE, FOR USE IN APPLYING THERAPEUTIC DEEP HEAT
Manufacturer (Section D)
DJO LLC
5919 sea otter place
ste 200
carlsbad CA 92010
Manufacturer (Section G)
DJO, LLC
5919 sea otter place
ste 200
carlsbad CA 92010
Manufacturer Contact
jim pomeroy
5919 sea otter place
ste 200
carlsbad, CA 92010
5128346301
MDR Report Key16449097
Report Number9616086-2023-00006
Device Sequence Number5530811
Product Code IMG
UDI-Device Identifier00888912021883
UDI-Public00888912021883
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K040285
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number2738
Device Catalogue Number2738
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2023
Initial Report FDA Received Date02/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-