It was reported that during a cryo ablation procedure, when the mapping catheter was connected and inserted into the patient, the mapping catheters seventh and eight electrode were experiencing noise.The cable was replaced without resolution.The electrodes seven and eight were removed to resolve the issue.The case was completed with cryo. no patient complications have been reported as a result of this event.
|
|
Product event summary: the 990063-020 mapping catheter with lot 225009083 was returned and analyzed.Visual inspection was performed, and a broken electrode wire was observed at the weld and a kink/twist was observed on the catheter shaft.Visual inspection of the shaft segment area showed the shaft was kinked approximately two inches from the lemo connector.The pin-to-pin continuity test failed due to an open circuit at electrode eight.Dissection of the suspected electrode related to the noise revealed the electrode wire was broken from the welding at electrode eight.Dissection of lemo connector assembly showed the lemo connector wire was intact with no issues.In conclusion, the mapping catheter failed the returned product inspection due to a broken electrode wire and a shaft kink/twist.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|