| Model Number |
PCB00 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Telephone number: (b)(6).The device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.Based on the manufacturing records review, and historical complaint review, there is no indication of a product malfunction or product quality deficiency.No nonconformity report, documentation or labeling changes, and escalations are required.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that haptic detachment was observed on the pcb00 model of intraocular lens (iol).Upon follow-up, it was confirmed that there was no patient involvement.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d9: device available for evaluation: yes.Date returned to manufacturer: mar 28, 2023.Section h3: device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection revealed that the complaint lens was received cut in half.The lens was cleaned and, no further issues were identified.Visual inspection of the complaint handpiece revealed that the plunger rod was fully advanced and in the locked position.The handpiece was disassembled and inspected and no issues that could cause or contribute to the complaint or observed issues were identified.Dimensional inspection was performed on the remaining haptic confirming that the haptic width and haptic thickness were manufactured within specification.Conclusion: based on the complaint investigation results, the product was released within specifications and a relationship between the device and the reported incident could not be determined.There is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|